Product details
Product introduction
Operating Conditions
Power Supply: 110V-240V, 50Hz/60Hz, 3A
Temperature: 2℃ to 40℃ (ideal: 20℃±5℃)
Humidity: ≤80% (storage: ≤93%)
Environment: No vibration, airflow, conductive dust, explosive, or corrosive gases
Installation Requirements
Stable Surface: Place on a solid, level table (avoid sliding surfaces to prevent vibration).
Ventilation: Maintain a 10 cm gap on both sides and 30 cm at the back for cooling.
Protection: Keep away from heat sources, water, and direct sunlight to prevent malfunctions.
Speed calculation formula
The centrifuge rotor can separate samples with a density ≤2.0 g/mL. For samples exceeding this density, calculate the allowable speed using the formula:
Allowable Speed (rpm) = Maximum Speed × √(2.0 g/mL ÷ Sample Density g/mL)
Calculation of RCF
Relative centrifugal acceleration (RCF) can be calculated by the formula:
RCF=1.118×r×n2×10-5
r-rotation radius, unit: cm; n—rotation speed, unit: rpm
Product Application
Mainly used for separation of blood or urine samples
Product Advantage
- Easy operation with an intuitive menu and easy-to-read display
- Speed can be set in RPM/RCF
- Maintenance-free brushless DC motor can quickly reach the target speed Quick spin function with press and hold pulse key.
- The device features a lid equipped with an electric lock and an automatic release mechanism, ensuring the lid unlocks automatically upon completion of operation.
- For the A6-50P rotor, braking time 2 steps adjustable
- A high strength plastic rotor gives excellent balance to guarantee stable working.
- Dual protection housing particularly benefitsfor safe running, prevent splashing or ejection.
- Gentle braking at low speeds with efficient separation